Calibration and Certification Services...

As an ISO-9001 Registered Company with ISO 17025 Accredited Procedures, we are committed to the highest quality calibration services with exceptional performance. Your CMM will be thoroughly cleaned, calibrated, adjusted and certified by one of our highly trained CMM technicians.

Standard Calibration Services

Your CMM will be cleaned, calibrated and certified using a High Precision Step Gage and Ballbar.

 

 

High Accuracy Laser Calibration Services

Your CMM will be cleaned, calibrated and certified using a High Accuracy Renishaw Laser and Ballbar.

All calibration procedures and accuracy verifications are in compliance with ASME B89.4.1-2000, ANSI / NCSL Z540-1, ISO 10012-1, ISO-9001:2008, ISO-17025 and/or the original manufacturers' specifications.  All of our accuracy standards are calibrated, certified and fully traceable to the National Institute of Standards and Technology (NIST).

Our Cleaning, Calibration and Certification Services will help you pass your stringent QS and ISO audits by providing you with the required before and after documentation pertaining the calibration & certification of your CMM.


Did you know???

Yearly Cleaning, Calibration & Certification of your CMM is highly recommended to insure its accuracy and performance.

Re-calibration is also required if your CMM has been moved, retrofitted, upgraded or if certain repairs have been made. Even a simple software update can effect the accuracy of your CMM.

If you have any questions or concerns about the accuracy of your CMM, please contact us and we will assist you to determine if your accuracy has been compromised in any way.


Calibration vs Certification...

Most CMM owners do not know the difference between Calibration & Certification and consider them to be the same. They are in fact two totally different procedures.

Calibration is to verify and correct accuracy errors from a known standard.
Certification is just the verification of accuracy errors from a known standard.

Just about anyone can attest to errors on a CMM, but it takes knowledge, skill, experience and the right equipment to find and correct all of the geometric errors that exist in your CMM's measuring volume. These errors include the Linear Accuracy, Squareness, Pitch, Roll, Yaw, Straightness, Repeatability and Volumetric Accuracy.

To perform a complete Calibration and Certification, your CMM must be evaluated according to the OEM specifications. After all of the data is collected, many mechanical, electronic and intricate error compensation adjustments are required to adjust it to within the manufacturers' specifications. We can insure the accuracy and performance of your CMM with our complete Cleaning, Calibration, Certification and Preventive Maintenance Services.